36
ALARMS, ADVISORIES AND PROMPTS
Advisories
Check Air Sensor
Infusion Complete VR =0
Low Battery
Channel Not In Use
Two beeps, slow flashing red light on infusing channel’s channel key; infusion continues.
Alarm
Meaning
Response
At installation of cassette:
a) air is detected in tubing;
b) tubing collar is not properly
seated;
or
c) air sensor is dirty or damaged.
VR has counted down to zero.
Channel is infusing at KVO rate.
30 minutes or less battery power
remaining.
Tw o m i n u t e s h a v e elapsed
since cassette was installed or
infusion was stopped.
Verify tubing collar is fully seated
in air sensor recess.
Verify tubing in air sensor recess
is not damaged, twisted or dirty.
Press
C
Cl
lr
rA
Ai
ir
r
on channel's Alarm
Information page. Three beeps
indicate air bubble is no longer in
front of air sensor.
If air is still present, remove
cassette and manually clear air
according to hospital policy.
If no air is present, clean air
sensor recess as directed in
cleaning instructions.
Enter new VR or, if same volume
is desired, press
R
RE
EP
PE
EA
AT
T
.
Press
E
En
nt
te
er
r
.
Press to resume primary
infusion rate.
Verify fluid flow.
Connect AC adapter power cord
to pump.
Plug into wall outlet.
Press
S
ST
TN
ND
DB
BY
Y
to place channel on
Standby,
OR
Press to start infusion,
OR
Remove cassette.
Summary of Contents for MedSystem III
Page 37: ...34 GETTING STARTED T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 45: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 42 ALARMS ADVISORIES AND PROMPTS...
Page 61: ...58 MAINTENANCE T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Page 63: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 60 MAINTENANCE...