Secondary Mode
Preparing the Administration Set and Container
This option allows two different rates of infusion to be
administered sequentially. When secondary volume remaining
reaches zero, primary infusion resumes automatically.
To avoid the possibility of concurrent flow during secondary
delivery of intermittent medications, set up the administration
set as recommended below.
B
C
A
START
STOP
MORE
OPTIONS
STANDARD
DISPLAY
O N / O F F
®
®
DLE
®
8"
• For Needle-Free sets, attach secondary to upper primary ‘Y’
site, below a check valve.
• Prepare the secondary IV container according to your
institution´s policy.
• Suspend secondary solution container at least 8 inches above
primary solution container.
• Press , or to select channel.
Setting a secondary rate over
275 ml/h may result in
concurrent flow with the
primary container.
WARNING