GASGUARD
®
TE10 Distribution System Manual
Air Products and Chemicals, Inc.
Page 3 - 10
The tables below list the exhaust requirement for the GASGUARD 500VMB
enclosure to meet UFC and NFPA code requirements.
Table 1
Minimum Exhaust Requirements for
GASGUARD 500 VMB Enclosures Containing Silane
ACCESS HATCH CLOSED [2]
ACCESS HATCH OPEN [3]
Enclosure
Size [1]
exhaust
flow
(cfm)
static
pressure
(in H20)
velocity
pressure
(in H20)
exhaust
flow
(m3/m)
static
pressure
(Pa)
velocity
pressure
(Pa)
exhaust
flow
(cfm)
velocity
pressure
(in H20)
exhaust
flow
(m3/m)
velocity
pressure
(Pa)
GG500
24” wide
375 -0.4 -0.23 10.6 -100 -57 400 -0.26 11.3 -65
GG500
30” wide
450 -0.5 -0.33 12.7 -124 -82 470 -0.36 13.3 -90
Notes
: [1] All vent stack duct sizes are 6 inch (154 mm) diameter.
[2] “Access Hatch Closed” values have been measured at the exhaust stack duct and
provide a minimum air velocity of 200 feet per minute (61 meters per minute) across
cylinder connections and panel fittings.
[
3] “Access Hatch Open” values have been measured while observing a minimum
velocity of 200 feet per minute (61 meters per minute) across the access hatch
opening.
[4] These recommended values assume that a 0.25 mm [0.010 inch] diameter
restrictive flow orifice (RFO) is installed in the Silane source cylinder valve per US
Code requirement
Table 2
Exhaust Recommendations for Gasguard Equipment Containing
Gases with a TLV
≥
0.20 ppm
(25% TLV
>
0.050 ppm)
Notes
:
[1
] ERC (Expected Release Concentration) values are taken from actual test data.
[2] ESI (Ex
hau
st Stack Insert) may have been specified by the customer to enhance
the readability of the exhaust pressure monitor. Refer to the system’s
Specification Sheet to determine if this option is used.
[3] These exhaust flow values are applicable to the 30 inch and 24 inch standard flow
Gasguard VMB enclosures.
[4] All standard Gasguard enclosures are sized for a 6 inch outer diameter (OD) duct
connection to the vent stack.
[5] “Access Hatch Closed” values have been measured, or calculated at the exhaust
stack duct.
[6] The actual values presented in "Access Port Open" are from experiments done on
a two-cylinder MC source system enclosure. Due to the similar design and smaller
access port of the GG500 VMB enclosure, balancing to Tables 2 and 3 will satisfy
meet or exceed UFC 80 face velocity requirements.
Maximum
ERC
[1]
Exhaust
Flow
Static
Pressure
Velocity
Pressure
Exhaust
Flow
Velocity
Pressure
Average
Duct
Velocity
Minimum
Face
Velocity
Average
Face
Velocity
Baffle
ppm
scfm
inches
H2O
inches
H2O
scfm
inches
H2O
fpm
fpm
fpm
No ESI
[2]
0.029
75
-0.5
0.05
318
0.255
1622
180
220
None
ESI
< 0.029
75
-1.5
0.15
237
0.775
1210
210
236
33%
Access Port Closed
Access Port Open
Summary of Contents for GASGUARD TE10
Page 2: ......
Page 12: ...GASGUARD TE10 Distribution System Manual Air Products and Chemicals Inc Page i x...
Page 70: ...GASGUARD TE10 Distribution System Manual Air Products and Chemicals Inc Page 3 4...
Page 92: ...GASGUARD TE10 Distribution System Manual Air Products and Chemicals Inc Page 4 12...
Page 94: ......
Page 154: ......
Page 166: ......
Page 174: ......
Page 180: ......
Page 188: ......
Page 190: ...GASGUARD TE10 Distribution System Manual Air Products and Chemicals Inc Page B 2...
Page 191: ...Appendix B Revision November 2004 Page B 3...
Page 192: ...GASGUARD TE10 Distribution System Manual Air Products and Chemicals Inc Page B 4...
Page 193: ...Appendix B Revision November 2004 Page B 5...
Page 194: ...GASGUARD TE10 Distribution System Manual Air Products and Chemicals Inc Page B 6...