Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Troubleshooting
Troubleshooting Assembly Level Problems
1-35
7.
Refer to
. Use a voltmeter with a small probe to measure A6J5-1; it should be
∼
12V.
•
If A6J5-1 is
∼
12V:
— Measure the connections in
— If the voltages are not present, replace the A6 DC-AC Inverter Interface Board.
— If the voltages are present, replace the LCD Display assembly.
•
If A6J5-1 is not
∼
12V, measure A6J1-3 or A6J1-4.
— If J1-3 or J1-4 is
∼
12V, replace the A6 DC-AC Inverter Interface and LCD assemblies.
— If J1-3 or J1-4 is not
∼
12V, go to
8.
Refer to
. Using the voltmeter measure A3J6-3.
•
If A3J6-3 is
∼
12V replace the W1 ribbon cable.
•
If A3J6-3 is not
∼
12V go to
Figure 1-11 A3J6-3 and A3J10-B16
Table 1-4
Connector
Voltage
J5-3
+11V
J5-4 (pwm_pic)
>+1.3V
A3J6-3
A3J10-B16
Front Panel
Rear Panel
Rear Panel
Front Panel
深圳德标仪器