
Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Troubleshooting
Self Tests Do Not Run or Self Tests Run Slowly
1-19
On the MXG ATE, if you are unable to establish a LAN connection, cycle power on the MXG ATE, press the LAN Reset hardkey.
This resets the MXG ATE to a factory preset where the Web-Enabled MXG ATE feature is available (i.e. Web Server softkey is set
to On).
For MXG ATEs, press the front panel LAN Preset key and wait for the front panel green LAN indicator light to stop blinking. It is
possible the hostname may have been changed from its default value. The MXG ATE hostname is not changed by pressing the
LAN Reset key. For information on using the MXG's USB port to query its IP address, refer to the Agilent Connectivity Guide
(E2094-90009) or to the LAN Connectivity FAQs for details on using the instrument over LAN.
For more information on LAN Connectivity, refer to the Agilent Connectivity Guide (E2094-90009), or the signal generator’s
Programming Guide, or to the LAN Connectivity FAQs for details on using the instrument over LAN.
1.
Open an internet browser.
2.
Enter the MXG’s hostname or IP address.
3.
Click
Signal Generator Web Control
from the menu on the left side of the web page to access the front panel Web-Enabled
MXG.
4.
Run a complete self test, go to
“Running Self Test” on page 1-16
.
Figure 1-3 Web-Enabled MXG
Self Tests Do Not Run or Self Tests Run Slowly
If self tests will not run or run slowly, the most likely cause is a faulty A1 Power Supply or A5 CPU assembly. The A5 CPU assembly relies
on correct power supply voltages. If the A1 Power Supply is faulty, it could cause a false A5 CPU assembly error.
1.
Troubleshoot the A1 Power Supply. Refer to
“A1 Power Supply” on page 1-26
2.
If troubleshooting the A1 Power Supply does not correct the problem, troubleshoot the A5 CPU assembly. Refer to
3.
If self tests still will not run or run slowly, the fault can be identified within the boot-up routine. Use the following
procedure to identify and isolate the problem.
Note: If you do
not
see this window, check to see if the window is hidden behind your browser
window or your web browser settings are set to block pop-ups. To use this feature, you need to
set your web browser to allow pop-ups for your instrument’s IP address.
Remote SCPI commands requires the Telnet feature on the computer. The Telnet feature is
available from a variety of sources. Some software updates can block (break) this Telnet
connection (e.g. Internet Explorer 7). When using Internet Explorer as a browser, only versions
<Internet Explorer 7 enable the Web-Enabled MXG SCPI feature.
To operate the signal generator, click the keys.
The Agilent MXG supports LXI Class B*
functionality. For more information on the LXI
standards, refer to
http://www.lxistandard.org/home
.
*LXI Class B Compliance testing using IEEE
1588-2008 not available at release.
If the SCPI Telnet softkey is not active, to display the SCPI Telnet box:
1) On the PC, click
Start > Run
2) In the dialogue box type:
Telnet xxx.xxx.xxx.xx [
IP address
of MXG
] 5024
[
This is the port number for connecting to the MXG
.]
Note: Telnet port 5023 is available for backwards compatibility (i.e. for
firmware versions <A.01.50).
深圳德标仪器
135-1095-0799