Agilent Technologies N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Post-Repair Procedures and Performance Verification
Leveled Output Power Verification (N5161A/62A
1
/81A/82A Only)
4-13
Leveled Output Power Verification (N5161A
1
This test verifies that the output power from the signal generator is within defined limits. The following table lists the preferred equipment
for this test.
•
/81A Output Power Test Procedure” on page 4-13
•
“N5162A/82A Output Power Test Procedure” on page 4-15
N5161A
/81A Output Power Test Procedure
1.
Zero and calibrate the power sensor to the power meter:
2.
Connect the equipment as shown:
3.
Preset the signal generator: Press
Preset
.
4.
Turn RF on: Press
RF On/Off
so that the RF On/Off LED lights.
5.
Turn modulation off: Press
Mod On/Off
so that the Mod On/Off LED turns off.
6.
Set the frequency: Press
Frequency
and enter the first frequency value listed in
1
The N5161A/62A performs the identical internal tests as in the MXG via a LAN connection to a Web Server. The front panel displays all hardkeys, softkeys and test results
as depicted in the illustration. For information on Web-Enabled Control refer to
“Accessing the MXG Web-Enabled Page” on page 1-18
and to the
Programming
Guide.
Test Equipment
Recommended Model
Power Meter
Agilent E4418B or E4419A/B E-Series power meter
Power Sensor,
Input: Type-N (m)
Agilent E9304A
Table 4-10Leveled Output Power Limits
Frequency
Limits (dB)
125 MHz
< 0.6dB
275 MHz
< 0.6dB
338 MHz
< 0.6dB
425 MHz
< 0.6dB