Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Troubleshooting
Blank or Missing Pixels (N5181A/82A/83A)
1-62
7.
Refer to
. Measure A6J1-3 and A6J1-4; they should have a voltage of +12 Vdc.
•
If A6J1-3 and A6J1-4 have a voltage of +12 Vdc, replace the A6 AC-DC Inverter Interface Board.
•
If A6J1-3 and A6J1-4 do not have a voltage of +12 Vdc, go to
8.
Refer to
. Measure A3J6-3 and A3J6-4; they should have a voltage of +12 Vdc.
•
If A3J6-3 and A3J6-4 have a voltage of +12 Vdc, replace the W1 ribbon cable.
•
If A3J6-3 and A3J6-4 do not have a voltage of +12 Vdc, go to
9.
Refer to
. Measure A3J10-B16; it should have a voltage of +12 Vdc.
•
If A3J10-B16 has a voltage of +12 Vdc:
a.
Replace the A3 RF assembly.
b.
For N5183A instruments, go to
Chapter 4, "Post-Repair Procedures and Performance Verification"
.
•
If A3J10-B16 does not have a voltage of +3 Vdc, go to
“A1 Power Supply” on page 1-26
to troubleshoot the A1 Power Supply.
10. Refer to
. Using an oscilloscope, turn the RPG knob and measure A6J2-26 and A6J2-30. There should be a 4V square pulse.
If there is no 4V square pulse, replace the Keypad assembly.
Blank or Missing Pixels (N5181A/82A/83A)
Verify functionality by performing the front panel LCD Pixel tests.
Press:
Utility
>
Instrument Info
>
Front Panel Tests
>
Pixel Screen Tests
•
If the same area of the display is blank during each pixel test, replace the LCD Display.
•
If an area of the display is only blank during some of the tests, replace the A6 DC-AC Inverter Interface board.
The Screen Saver is Not Functioning (N5181A/82A/83A)
The screen saver delay can be set from 1 to 12 hours. The PIC Controller controls the screen saver by turning the LCD backlight on or off.
1.
Verify the signal generator is turned on.
2.
Set the delay time to 1 hour. Press:
Utility
>
Display
>
Screen Saver Delay
, then turn the RPG knob.
3.
Allow the screen saver delay time to elapse.
4.
Check to see if the display has gone dark.
5.
If the display has not gone dark, replace the A6 DC-AC Inverter Interface assembly.
6.
Set the delay time to 1 hour.
7.
Allow the screen saver delay time to elapse.
8.
Check to see if the display has gone dark.
9.
If the display still has not gone dark, replace the A5 CPU assembly.
Boot Time Diagnostics Mode (N5161A/62A Only)
CAUTION
This section should only be used by qualified service personnel.
This procedure serves as a full test of the front panel on an MXG ATE, as the LEDs and LAN Reset key must be functioning properly in
order for the instrument to proceed to the next mode.
To run the boot time diagnostics mode:
1.
Turn the instrument off.
2.
Press
and hold
the LAN Reset hardkey then press
and release
the power switch, until the 3 front panel LEDs start blinking. They will
blink for approximately 10 seconds.