Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Assembly Replacement
Power Switch
3-21
Power Switch
Tools Required
•
T-10 driver
Removal Procedure
for this procedure.
1.
Disconnect the power cord.
2.
Remove the covers from the signal generator. Refer to
.
3.
Remove the front-panel from the signal generator. Refer to
.
4.
Remove the A6 DC-AC Inverter Interface Board from the signal generator. Refer to
”A6 DC-AC Inverter Interface Board
(N5181A/82A/83A only)” on page 3-16
5.
Remove the LCD Display from the signal generator. Refer to
”LCD Display (N5181A/82A/83A only)” on page 3-14
.
6.
Using the T-10 driver, remove the two screws (1) that secure the Power Switch (2) to the front-panel frame.
7.
Remove the Power Switch.
Replacement Procedure
1.
Reverse the order of the removal procedure.
2.
Torque all screws to 9 in–lbs.
3.
Perform the post-repair performance tests that pertain to this replacement procedure, see
Chapter 4, Post-Repair Procedures and
.
Figure 3-10Power Switch
1
2
screws not shown
Note: N5161A/62A use the longer power switch cable