Hardware Information
10
Setting the 8-bit Configuration Switch
Legend:
0 (switch down), 1 (switch up), x (any position)
N O T E
To perform any LAN configuration, SW1 and SW2 must be set to OFF. For details on the
LAN settings/configuration refer to chapter LAN Configuration.
Table 7
8-bit Configuration Switch (with on-board LAN)
Mode
Function
SW 1
SW 2
SW 3
SW 4
SW 5
SW 6
SW 7
SW 8
LAN
0
0
Link Configuration
Init Mode Selection
Auto-negotiation
0
x
x
x
x
x
10 MBit, half-duplex
1
0
0
x
x
x
10 MBit, full-duplex
1
0
1
x
x
x
100 MBit, half-duplex
1
1
0
x
x
x
100 MBit, full-duplex
1
1
1
x
x
x
Bootp
x
x
x
0
0
0
Bootp & Store
x
x
x
0
0
1
Using Stored
x
x
x
0
1
0
DHCP
x
x
x
1
0
0
Using Default
x
x
x
0
1
1
TEST
1
1
System
NVRAM
Boot Resident System
1
x
Revert to Default Data (Coldstart)
x
x
x
1
N O T E
When selecting the mode TEST, the LAN settings are: Auto-Negotiation & Using Stored.
N O T E
For explanation of "Boot Resident System" and "Revert to Default Data (Coldstart)" refer to
Summary of Contents for G7104A
Page 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Page 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Page 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Page 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Page 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Page 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...