Agilent 1290 Infinity II Flexible Pump User Manual
7
Maintenance
Release a Stuck Inlet Valve
Release a Stuck Inlet Valve
Tools required
p/n
Description
9301-0411
Syringe, Plastic
0100-1681
Syringe adapter luer/barb
0100-1710
Mounting Tool for Tubing Connections
Beaker
C A U T I O N
Pressure damages the multi-channel gradient valve (MCGV) and/or degasser
➔
Never apply pressure to the MCGV or degasser.
➔
Directly connect the syringe to the inlet weaver.
1
Remove tubing connections channels A, B, C and D to the
MCGV such that you can access the inlet weaver.
2
Open the fitting at the center of the multi-channel
gradient valve (MCGV). Do not open the screw marked
with the red cross. Remove the inlet weaver from the
MCGV.
Fitting