Agilent 1290 Infinity II Flexible Pump User Manual
209
LAN Configuration
11
Automatic configuration with Bootp
Changing the IP Address of an Instrument Using the Agilent BootP
Service
Agilent BootP Service starts automatically when your PC reboots. To
change Agilent BootP Service settings, you must stop the service, make the
changes, and then restart the service.
Stopping the Agilent BootP Service
1
From the Windows control panel, select
Administrative Tools > Services
. The
Services
screen appears.
Figure 35
Windows Services screen
2
Right- click
Agilent BootP Service
.
3
Select
Stop
.
4
Close the
Services and Administrative Tools
screen.
Editing the IP address and other parameters in EditBootPSettings
1
Select
Start > All Programs > Agilent BootP Service
and select
Edit BootP
Settings
. The
BootP Settings
screen appears.
2
When the
BootP Settings
screen is first opened, it shows the default
settings from installation.
Summary of Contents for G7104A
Page 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Page 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Page 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Page 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Page 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Page 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...