Agilent 1290 Infinity II Flexible Pump User Manual
137
Maintenance
7
Replace the High Pressure Outlet Filter or Filter Frit
3
Unscrew the high pressure outlet filter.
4
Replace the filter frit.
5
Mount the high pressure outlet filter using a torque
wrench set to 16 Nm.
6
Reinstall the high pressure outlet filter to the filter holder.
7
Mount the capillary connection to the pressure sensor (1)
and to the Multipurpose valve (2).
;g^i
'#
&#