Agilent 1290 Infinity II Flexible Pump User Manual
7
Maintenance
Pump Head Procedures
C A U T I O N
Damage to the pump head
Using a wrong torque will damage the pump head.
➔
For handling the torque wrench, setting and applying
the right torque, consult the manual of your torque
wrench.
11
For the primary pump head, install the inlet valve and
outlet valve using the torque wrench, which is included
to the 1290 Infinity Service Kit p/n 5067-4699. Set 10 Nm
for the inlet valve and 10 Nm for the outlet valve.
C A U T I O N
Damage to the pump head
Using a wrong torque will damage the pump head.
➔
For handling the torque wrench, setting and applying
the right torque, consult the manual of your torque
wrench.
12
For the secondary pump head, install the outlet filter
adapter using the torque wrench (14 mm hex bit), which
is included to the 1290 Infinity Service Kit p/n 5067-4699,
set to 16 Nm.
10 Nm
10 Nm
16 Nm
Summary of Contents for G7104A
Page 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Page 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Page 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Page 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Page 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Page 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...