![Agilent Technologies G7104A User Manual Download Page 117](http://html1.mh-extra.com/html/agilent-technologies/g7104a/g7104a_user-manual_2867748117.webp)
Agilent 1290 Infinity II Flexible Pump User Manual
117
Maintenance
7
Pump Head Procedures
Replacing the Heat Exchanger
Tools required
Description
Wrench, 19 mm
Parts required
p/n
Description
G4220-81013
Heat Exchanger (secondary pump head only)
G4220-20028
Headless screw for 1290 Infinity pump heads
G4220-20001
Spacer Fitting
Preparations
•
Remove the pump head assembly from the pump
•
Remove the secondary pump head from the link plate
C A U T I O N
Loss of small spacer fitting
Inside the secondary pump head is a small spacer fitting, which can be dropped easily
when removing the heat exchanger.
➔
The heat exchanger does not need to be removed for pump head maintenance.
1
Remove the 19 mm screw at the front of the secondary
pump head.
2
Remove the front plate.