Agilent 1290 Infinity II Flexible Pump User Manual
99
Maintenance
7
Replace the Seal Wash Pump
Replace the Seal Wash Pump
When
In case of wear of the seal wash pump
Parts required
p/n
Description
5065-4445
Peristaltic pump with Pharmed tubing
5065-9978
Tubing, 1 mm i.d., 3 mm o.d., silicone, 5 m
Preparations
Remove the flow connections from and to the seal wash pump.
1
For removing the seal wash pump, press the clips (1) and
pull the pump to the front (2).
2
Insert the pump clips to the holes in the module housing.
3
Fix the seal wash tubings to the peristaltic pump inlet (1)
and the from the peristaltic pump outlet to the primary
pump head inlet (2).