![Agilent Technologies G7104A User Manual Download Page 68](http://html1.mh-extra.com/html/agilent-technologies/g7104a/g7104a_user-manual_2867748068.webp)
68
Agilent 1290 Infinity II Flexible Pump User Manual
6
Error Information
Pump Error Messages
Pump drive encoder defect
Error ID: 29209
Defect pump drive encoder.
•
Parameter: 1 – 2 referring to pump drive
Multi Purpose Valve failed
Error ID: 29231
Lost steps of the purge valve encoder.
•
Parameter: None
Probable cause
Suggested actions
1
Defect encoder.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Multi purpose valve drive mechanically
blocked or defect.
•
Check installation of multi purpose valve
head.
•
Replace multi purpose valve head.
Summary of Contents for G7104A
Page 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Page 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Page 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Page 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Page 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Page 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...