![Agilent Technologies G7104A User Manual Download Page 208](http://html1.mh-extra.com/html/agilent-technologies/g7104a/g7104a_user-manual_2867748208.webp)
208
Agilent 1290 Infinity II Flexible Pump User Manual
11
LAN Configuration
Automatic configuration with Bootp
5
Make these entries for the instrument:
•
MAC address
•
Host name, Enter a Hostname of your choice.
The Host Name must begin with "alpha" characters (i.e. LC1260)
•
IP address
•
Comment (optional)
•
Subnet mask
•
Gateway address (optional)
The configuration information entered is saved in the Tab File.
6
Click
OK
.
7
Leave
Edit BootP Addresses
by pressing
Close
.
8
Exit
BootP Settings
by pressing
OK
.
9
After each modification of the BootP settings (i.e. EditBootPSettings) a
stop or start of the BootP service is required for the BootP service to
accept changes. See
“Stopping the Agilent BootP Service”
or
“Restarting the Agilent BootP Service”
10
Power cycle the Instrument.
OR
If you changed the IP address, power cycle the instrument for the
changes to take effect.
11
Use the PING utility to verify connectivity by opening a command
window and typing:
Ping 192.168.254.11 for example.
The Tab File is located at
C:\Documents and Settings\All Users\Application Data\Agilent\BootP\TabFile
Summary of Contents for G7104A
Page 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Page 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Page 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Page 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Page 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Page 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...