![Agilent Technologies G7104A User Manual Download Page 186](http://html1.mh-extra.com/html/agilent-technologies/g7104a/g7104a_user-manual_2867748186.webp)
186
Agilent 1290 Infinity II Flexible Pump User Manual
10
Hardware Information
Setting the 8-bit Configuration Switch
Special Settings
The special settings are required for specific actions (normally in a service case).
Boot-Resident
Firmware update procedures may require this mode in case of firmware
loading errors (main firmware part).
If you use the following switch settings and power the instrument up
again, the instrument firmware stays in the resident mode. It is not
operable as a module. It only uses basic functions of the operating system
for example, for communication. In this mode the main firmware can be
loaded (using update utilities).
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode
with default parameter settings.
If you use the following switch settings and power the instrument up
again, a forced cold start has been completed.
N O T E
The tables include both settings for modules – with on-board LAN and without on-board
LAN. They are identified as LAN and no LAN.
Table 8
Boot Resident Settings (On-board LAN)
Mode Select
SW1
SW2
SW3
SW4
SW5
SW6
SW7
SW8
TEST/BOOT
1
1
1
0
0
0
0
0
C A U T I O N
Loss of data
Forced cold start erases all methods and data stored in the non-volatile memory. Exceptions
are calibration settings, diagnosis and repair log books which will not be erased.
➔
Save your methods and data before executing a forced cold start.
Table 9
Forced Cold Start Settings (On-board LAN)
Mode Select
SW1
SW2
SW3
SW4
SW5
SW6
SW7
SW8
TEST/BOOT
1
1
0
0
0
0
0
1
Summary of Contents for G7104A
Page 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Page 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Page 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Page 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Page 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Page 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...