Chapter 1
85
Instrument Messages
Instrument Messages
In
str
u
me
n
t Me
s
s
a
g
es
10901 to 10925: NADC Error Messages (Option BAE)
An error number in this range
indicates the instrument has detected an error relating to the NADC personality.
10902
Sync word not found
In an EVM measurement, the sync word is not found and the
synchronization cannot be established when
Sync Word
is selected in the
Burst Sync
menu.
10904
Valid NADC burst not found
A valid NADC burst is not found when the
Device
is MS.
10906
Signal too noisy
The valid EVM measurement cannot be performed, because the input
signal is too noisy.
10908
Burst Delay out of limit for EVM (2 ms)
In an EVM measurement, the
Burst Delay
value must be less than 2 ms.
10926 to 10950: PDC Error Messages (Option BAE)
An error number in this range
indicates the instrument has detected an error relating to the PDC personality.
10926
Sync word not found
In an EVM measurement, the sync word is not found and the
synchronization cannot be established when
Sync Word
is selected in the
Burst Sync
menu.
10928
Valid PDC burst not found
A valid PDC burst is not found when the
Device
is MS.
10930
Signal too noisy
The valid EVM measurement cannot be performed, because the input
signal is too noisy.
10932
Burst Delay out of limit for EVM (2 ms)
In an EVM measurement, the
Burst Delay
value must be less than 2 ms.
10951 to 10975: cdmaOne Error Messages (Option BAC)
An error number in this
range indicates the instrument has detected an error relating to the cdmaOne personality.
10952
Signal exceeds maximum allowable power
10954
Input overload
Excessive input power has been detected which will cause the ADC to clip
the signal. Reduce the signal level, change the attenuator/max total power
setting (under
Input
menu), or press
Restart
if the
RF Input Range
is
Auto
.
Summary of Contents for ESA Series
Page 6: ...6 Contents Table of Contents ...
Page 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...