48
Chapter 1
Instrument Messages
Instrument Messages
Ins
tr
u
me
nt Messa
g
e
s
197
Marker Freq Count HW Failure
201
Option not installed
The desired operation cannot be performed because a required option is
not installed. For example, pressing
Source
with no tracking generator
installed in the ESA analyzer, or issuing a
POW:GAIN
command without
Option 1DS or 110 installed in the PSA analyzer will generate this error.
202
No peak found
No signal peak was found that meets the criteria under
Peak Search
,
Search Param Criteria
.
204 ESA only
TG Frequency Limit
The tracking generator has reached the limit of its allowable frequency
range.
205
Ext Ref
205
Command not recognized
Indicates that the command sent from the remote interface was not
recognized. Check the programming guide for correct syntax.
206
Unable to initialize flatness data
A failure occurred in setting the flatness data in the internal EEROM. Get
in touch with your local Agilent Technologies sales and service office.
207
Unable to store flatness data
A failure occurred in setting the flatness data in the internal EEROM. Get
in touch with your local Agilent Technologies sales and service office.
208
Preselector is not available at this frequency.
A preselector is available, but the marker is at a frequency which does not
support preselection (~<3 GHz). Preselector centering can be performed
only for frequencies above 3.025 GHz.
209
Preselector centering failed
An attempt to center the preselector failed. You tried to center the
preselector with the analyzer span too wide. Preselector centering can be
performed only for spans below 1.123 GHz.
211
RBW limited to 1 kHz when Span > 5 MHz
In spans greater than 5 MHz, narrow (digital) resolution bandwidths,
below 1 kHz, are not available.
213
Span limited to 5 MHz when RBW < 1 kHz
In narrow (digital) resolution bandwidths, below 1 kHz, spans greater
than 5 MHz are not available.
Summary of Contents for ESA Series
Page 6: ...6 Contents Table of Contents ...
Page 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Page 33: ...33 Instrument Messages 1 Instrument Messages ...
Page 111: ...111 Functional Tests 2 Functional Tests ...
Page 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Page 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Page 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...