Chapter 1
65
Instrument Messages
Instrument Messages
In
str
u
me
n
t Me
s
s
a
g
es
10156
No Fast ADC hardware installed.
The analyzer cannot use sweep times of less than 5 msec when Option
B7D or Option AYX is not installed resulting in all the radio standards not
being supported.
10156
The calibration data is invalid, and has been cleared.
10157
Tracking Generator hardware is not present. Meas unavailable.
The measurement requires a built-in tracking generator.
10158
This operation requires a measurement to be active.
10159
Entire trace is below threshold level
The measurement cannot operate properly because the trace has
completely fallen below the threshold level. Change the threshold level to
view trace.
10160
Valid burst not found
Cannot find a valid burst, either because it is not fully on the display, or
the level is too low to be detected properly.
10160
Upper Custom Mask is Invalid!
The user-specified upper custom mask cannot be resolved into a limit line.
The format may be bad.
10161
No Fast ADC hardware installed.
Cannot use sweep times less than 5 ms, due to the lack of a fadc, therefore
the meas will be severely restricted.
10161
Lower Custom Mask is Invalid!
The user-specified lower custom mask cannot be resolved into a limit line.
The format may be bad.
10162
Resolution BW <300 kHz.
This error message is a warning that the resolution bandwidth has been
set below 300 kHz. The test results will not meet GSM specifications.
10163
Cannot find the Power vs Time Limits File.
The limit line definition file for the GSM standards has been deleted. This
message is displayed while the
Measure
key is grayed out. Reinstall the
GSM measurement personality.
10164
Band Measurement not defined for Out of Band.
You are attempting to monitor the band but have set the frequency outside
the band. Reset the band for the particular standard for which you are
testing or use the channel setting which does not require a frequency to be
set. (
Meas Setup
,
Method (Channel)
).
Summary of Contents for ESA Series
Page 6: ...6 Contents Table of Contents ...
Page 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Page 33: ...33 Instrument Messages 1 Instrument Messages ...
Page 111: ...111 Functional Tests 2 Functional Tests ...
Page 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Page 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Page 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...