Chapter 4
165
ESA Functional Tests
Frequency Readout Accuracy
ESA Fu
ncti
onal
T
ests
b. Press
System
,
Power On/Preset
,
Preset Type
(Factory),
Preset
on the analyzer, then
wait for the preset routine to finish. Set the analyzer by pressing the following keys:
Frequency
, Center Freq, 1.490,
GHz
SPAN
, 10,
MHz
BW/Avg
,
Res BW
, 100,
kHz
Video BW
, 30,
kHz
CAUTION
Use only 75
Ω
cables, connectors, or adapters on analyzers with 75
Ω
inputs,
or damage to connectors will occur.
3. Press
Peak Search
on the analyzer to measure the frequency readout accuracy. Record
this in the Marker Frequency Readout column in
.
4. Press
Span
, 100,
kHz
,
BW/Avg
,
Res BW
, 1,
kHz
,
Video BW
, 1,
kHz
.
5. Press
Peak Search
on the analyzer to measure the frequency readout accuracy. Record
this in the Marker Frequency Readout column in
.
NOTE
The Frequency Readout Accuracy in now complete. Continue with the Marker
Count Accuracy functional check.