Chapter 4
179
ESA Functional Tests
Resolution Bandwidth Switching Uncertainty
ESA Fu
ncti
onal
T
ests
Video BW
, 1,
kHz
2. Press
AMPLITUDE
and use the knob to adjust the reference level until the signal appears
five divisions (mid-screen) below the reference level. Press the following keys on the
analyzer:
Peak Search
Marker
,
Delta
FREQUENCY
,
Signal Track
(On)
3. Set the analyzer Resolution Bandwidth and Span according to
.
4. Press
Peak Search
, then record the
∆
Mkr 1 amplitude reading in
5. Repeat step 3 and 4 for each of the remaining resolution bandwidth and span settings
. The
∆
Mkr 1 amplitude reading should be within the range indicated
in the table
Table 4-7
Resolution Bandwidth Switching Uncertainty Worksheet
Resolution
Bandwidth
Setting
SPAN
Setting
∆
Mkr 1 Amplitude
Reading
1 kHz
50 kHz
0 (Ref)
3 kHz
50 kHz
10 kHz
50 kHz
30 kHz
500 kHz
100 kHz
500 kHz
300 kHz
5 MHz
1 MHz
10 MHz
3 MHz
10 MHz