66
Chapter 1
Instrument Messages
Instrument Messages
Ins
tr
u
me
nt Messa
g
e
s
10165
Carrier Not Present.
10166
Cannot update the list of cable types.
The cable file may have been moved or deleted accidentally. Reinstall the
measurement personality. This message is applicable to the GSM (Option
BAH) and Distance to Fault (Option 225) measurement personality
options.
10166
Cannot locate the file that contains the list of cable types.
The cable file may have been moved or deleted accidentally. Reinstall the
measurement personality. This message is applicable to the GSM (Option
BAH) and Distance to Fault (Option 225) measurement personality
options.
10167
The file containing the list of cable types has been updated.
10168
Cannot update the list of cable types on drive C:
The file update failed.
10170
The Cable Fault Measurement is active. Mode Setup is disabled.
Mode setup is not available in the cable fault utility.
10172
Sweep Time too fast (<2 sec)
The sweep time must be set to 2 seconds or longer for the results to be
valid.
10173
Carrier Not Present.
10177
There are no spurs to inspect.
You have attempted to switch the
Inspect Spur
menu key to the
On
position
after the measurement has finished, but found no spurs.
10178
Carrier Not Present.
10179
Carrier Present. Test Stopped!
A carrier was found in the transmit band. Either disable the carrier or
insert a bandpass filter for the receive bandwidth.
10180
Gate option not installed. Results may not be accurate.
This measurement method requires the use of the time-gate (option 1D6)
in order to gate the spectrum during the 50-90% part of the burst. If the
gate option is not installed, the measurement will still run although this
warning will be displayed.
10181
Upper Custom Mask is Invalid!
The user-specified upper custom mask cannot be resolved into a limit line.
The format may be bad.
Summary of Contents for ESA Series
Page 6: ...6 Contents Table of Contents ...
Page 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Page 33: ...33 Instrument Messages 1 Instrument Messages ...
Page 111: ...111 Functional Tests 2 Functional Tests ...
Page 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Page 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Page 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...