192
Chapter 4
ESA Functional Tests
Tracking Generator Level Flatness: E4402B, E4403B, E4404B, E4405B, E4407B and E4408B
E7402A, E7403A, E7404A, and E7405A, Option 1DN
ESA
Functional T
ests
6. Set the analyzer center frequency to 100 kHz by pressing
FREQUENCY
,
Center Freq
, 100,
kHz
.
7. Enter the appropriate power sensor Cal Factor for the test frequency into the power
meter as indicated on the label of the power sensor. This must be done at each test
frequency.
8. Record the power level displayed on the power meter as the Level Flatness in
.
9. Repeat steps 5 through 7 to measure the flatness at each center frequency setting listed
⇑
may be used to tune to center frequencies above 500 MHz.
10.Press
System
,
Alignments
,
Auto Align
,
On
.
Table 4-10 Tracking Generator Level Flatness Worksheet
Center Frequency
Level Flatness (dB)
100 kHz
5 MHz
40 MHz
50 MHz
0 (Ref)
80 MHz
500 MHz
1000 MHz
1500 MHz
2000 MHz
2300 MHz
2500 MHz
2700 MHz
3.0 GHz