Chapter 4
163
ESA Functional Tests
Frequency Readout Accuracy
ESA Fu
ncti
onal
T
ests
Frequency Readout Accuracy
Test Limits
Test Description
The frequency readout accuracy of the analyzer is tested with an input signal of known
frequency. Frequency reference error is eliminated by using the same frequency standard
for the analyzer and the synthesized sweeper.
Span
Minimum)
Maximum
10 MHz
1.48988 GHz
1.49012 GHz
100 kHz
1.4899988 GHz
1.4900012 GHz
Marker Count Accuracy with Counter
Resolution at 1 Hz
1.489999999 GHz
1.490000001 GHz
Item
Critical Specifications
(for this test)
Recommended
Agilent Model
Signal Sources
Synthesized Sweeper
10 MHz to 1.5 GHz
External Reference Input
8340A/B or 836XX
Series
Adapters
Type-N (f), to APC
3.5(f)
1250-1745
Cables
Type-N, 152-cm
(60-in)
11500D
BNC, 122-cm (48-in)
10503A
Additional Equipment for 75-Ohm Input
Pad, minimum loss
11852B
Type-N (f), to BNC
(m), 75
Ω
1250-1534