10417
This operation requires a measurement to be active.
The analyzer cannot perform this operation, because it requires a
measurement to be running and no measurement is running.
10418
Demod Format setting does not allow this value for
Points/Symbol.
Invalid Points/Symbol selection attempted for the given Demod Format
setting.
10419
Carrier Not Present.
A carrier signal/burst is expected at the analyzer input. This signal cannot
be found; however, the measurement will still run. Check input signal
connection.
10420
Valid burst not found.
This error only occurs when
RF Amptd
is selected for
Burst Sync
. It
indicates that a valid burst envelope which meets the specified Burst
Search Threshold was not found within the specified Burst Search Length.
Note that the demodulation measurement will proceed by attempting to
demodulate the signal without Burst Sync alignment. Possible causes may
be:
1. Carrier signal is not actually bursted.
2. Burst Search Threshold and/or Burst Search Length may need to be
adjusted.
10421
Cannot lock to carrier.
This error only occurs when attempting to demodulate OFFSET QPSK
signals. It indicates that the demodulation algorithm is unable to lock to
your signal. Possible causes may be:
1. There is no carrier signal.
2. Carrier signal is present, but Demod settings do not match the
modulation format of the OFFSET QPSK signal being measured.
3. OFFSET QPSK signal is too noisy to achieve carrier lock.
10425
B7D DSP Code Install Fail Opt.231. Contact your Service
Center.
The Digital Signal Processor algorithm files required to for the 89600 VSA
SW are not present in the analyzer. Reinstall the measurement
personality and contact your service center.
10426
Opt 231 DSP algorithm code file requires upgrade.
The Digital Signal Processing algorithm code file revision is not supported
by the currently loaded personality version. Refer to the user guide for the
personality in use for more information on installation/upgrade.
Summary of Contents for ESA Series
Page 6: ...6 Contents Table of Contents ...
Page 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Page 33: ...33 Instrument Messages 1 Instrument Messages ...
Page 111: ...111 Functional Tests 2 Functional Tests ...
Page 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Page 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Page 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...