186
Chapter 4
ESA Functional Tests
Second Harmonic Spurious Responses
ESA
Functional T
ests
6. Wait for two sweeps to finish, then press the following analyzer keys:
Peak Search
Mkr
→
Mkr
→
CF Step
Marker
,
Delta
FREQUENCY
7. Press the
⇑
key on the analyzer to step to the second harmonic (at 80 MHz). Press
Peak Search
. The marker delta amplitude reading should be less than the Maximum
value listed in the Test Limits Table.