Chapter 4
187
ESA Functional Tests
Tracking Generator Level Flatness: Models E4401B and E4411B E7401A, Options 1DN and
1DQ
ESA Fu
ncti
onal
T
ests
Tracking Generator Level Flatness: Models E4401B and
E4411B E7401A, Options 1DN and 1DQ
Test Limits
Test Description
A calibrated power sensor is connected to the tracking generator output to measure the
power level at 50 MHz. The power meter is set for REL mode so that future power level
readings are in dB relative to the power level at 50 MHz. The tracking generator is then
stepped to several frequencies throughout its range. The output power difference relative
to the power level at 50 MHz is measured at each frequency and recorded. Analyzers with
75
Ω
tracking generators are only tested from 1 MHz to 1500 MHz.
Maximum
Maximum
Flatness
≤
10 MHz
,
50
Ω
–2.5 dB
2.5 dB
Flatness > 10 MHz
,
50
Ω
–2.0 dB
2.0 dB
Flatness > 10 MHz, 75
Ω
–3.0 dB
3.0 dB
Flatness > 10 MHz, 75
Ω
–2.5 dB
2.5 dB
Item
Critical Specifications
(for this test)
Recommended
Agilent Model
Meters
Power Meter
438A or E4418A,
E4419A
RF Power Sensor
Frequency Range:
100 kHz to 1.5GHz
8482A
Adapters
Type-N (m to BNC (f)
(2)
1250-1476
Cables
BNC, 122-cm (48-in)
(2)
10503A
BNC, 23-cm (9-in)
10502A
Additional Equipment for 75
Ω
Input
75
Ω
Power Sensor
Frequency Range:
1 MHz to 1.5GHz
8483A
Type-N (f) to BNC
(m), 75
Ω
Adapter
1250-1534
Summary of Contents for ESA Series
Page 6: ...6 Contents Table of Contents ...
Page 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Page 33: ...33 Instrument Messages 1 Instrument Messages ...
Page 111: ...111 Functional Tests 2 Functional Tests ...
Page 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Page 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Page 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...