144
Chapter 5
Functional Testing
Frequency Response (Flatness)
Frequency Response (Flatness)
Test Limits
Signal amplitude should remain within
±
2 dB (4 dB peak-to-peak) from
10 MHz to 4 GHz.
Test Description
The frequency response test measures amplitude error as a function of
the tuned center frequency. Measurements are made at various points
from 10 MHz to 4 GHz. The signal source amplitude is measured with a
power meter to eliminate errors due to source flatness. The power
meter is zeroed and calibrated before starting the measurement.
Required Equipment
(include cables and connectors)