146
Chapter 5
Functional Testing
Frequency Response (Flatness)
5. Initialize the parameters for the instrument, as listed in Table 5-2,
“UUT Initialization,” by performing the following steps:
a. Press
MODE
,
Basic
.
b. To expand the Spectrum window to full screen, ensure that the
Spectrum window is highlighted and press
Zoom
.
If the Spectrum window is not highlighted, press
Next Window
until the Spectrum window is highlighted and then press
Zoom
.
c. Press
FREQUENCY
,
Center Freq
, 10, and
MHz
.
d. Press
Input
,
Input Atten
, 10, and
dB
.
e. Press
Meas Setup
,
Average
,
Avg Number
, 10, and
Enter
.
f. Press
View/Trace
,
Trace Display
, and
Average
.
g. Press
Marker
,
Trace
, and
Spectrum Avg
.
6. Preset the signal source and initialize the test equipment
parameters as listed in Table 5-3, “Signal Source Initialization.”
Table 5-2
UUT Initialization
Parameter
Setting
Mode
Basic
Measure
Spectrum
Frequency
Center
10 MHz
Input Attenuation
10 dB
Average
On/Off
On
Mode
Exp
Type
Log-Pwr Avg
Number
10
Trace Display
Average
Marker
Spectrum Avg
Table 5-3
Signal Source Initialization
Parameter
Setting
Frequency
10 MHz
Amplitude
−4
dBm
RF output
On
Summary of Contents for E4406A
Page 4: ...4 ...
Page 10: ...10 Contents ...
Page 138: ...138 Chapter4 Making Measurements Front Panel Test ...
Page 139: ...139 5 Functional Testing ...
Page 214: ...214 Chapter6 If You Have a Problem Returning Your Instrument to Agilent Technologies ...
Page 216: ...216 Chapter6 If You Have a Problem Repair Tag Goes Here Remove This Page ...
Page 217: ...217 7 Options and Accessories ...
Page 226: ...226 Chapter7 Options and Accessories Accessories ...