Chapter 4
135
Making Measurements
Sensors
Results
Window One (IF Signal Level)
Shows a stripchart of the IF signal amplitude in dBm.
Window Two (Cal Osc Level)
The Cal Oscillator window is not implemented.
Window Three (RF Temperature)
Shows a stripchart of the RF assembly temperature in Celsius.
Window Four (numeric results)
IF Signal Level: The ADC number for the detected 21.4 MHz IF
signal at the input to the AIF. Typical values with the input
attenuator set to 0 dB: 20 (no signal applied); 23(- 10 dBm); 27(- 5
dBm), and 38(0 dBm).
Cal Osc Level: Not implemented.
RF Temp: Shows current temperature in Celsius.