202
Chapter 6
If You Have a Problem
Error Message Descriptions
(5)
Analog IF sample rate osc unlocked
The 30 MHz sample rate oscillator on the A12 analog
IF assembly is unlocked.
(6)
Even second clock failing
The even second clock is unlocked.
(7)
No application file
(8)
Catalog incomplete
(9)
Application not licensed
License key “word” is not entered into instrument
memory.
(10)
Application not installed
Measurement application could not be found.
(11)
Invalid application file
Caused by an invalid personality file.
(12)
Application load failed
Measurement application could not load.
(13)
Invalid trace number
Caused by an invalid trace number. Some
measurements only have 1 or 2 valid traces, rather
then the indicated 4.
(14)
Trace data not ready
This may be caused by sending a command that asks
for trace data from a measurement that has not
finished calculating, or from a measurement that is not
currently active (running).
(15)
Measurement data not available
This may be caused by sending a command that asks
for data from a measurement that is not currently
active (running).