Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
407
:TRIGger:LIN:TRIGger
(see
Command Syntax
:TRIGger:LIN:TRIGger <condition>
<condition> ::= {SYNCbreak | ID}
The :TRIGger:LIN:TRIGger command sets the LIN trigger on condition to be
Sync Break (SYNCbreak) or Frame Id (ID).
Query Syntax
:TRIGger:LIN:TRIGger?
The :TRIGger:LIN:TRIGger? query returns the current LIN trigger value.
Return Format
<condition><NL>
<condition> ::= {SYNC | ID}
Errors
•
See Also
•
"Introduction to :TRIGger Commands"
•
•
":TRIGger:LIN:SIGNal:DEFinition"
•
N O T E
The ID option is available when the automotive CAN and LIN serial decode option (Option
AMS) has been licensed.