Commands Quick Reference
2
Agilent 6000 Series Oscilloscopes Programmer's Reference
39
Table 6
:CALibrate Commands Summary
Command
Query
Options and Query Returns
n/a
:CALibrate:DATE? (see
<return value> ::=
<day>,<month>,<year>; all in NR1
format
:CALibrate:LABel
<string> (see
)
:CALibrate:LABel? (see
<string> ::= quoted ASCII string up
to 32 characters
:CALibrate:STARt (see
n/a
n/a
n/a
:CALibrate:STATus? (see
<return value> ::=
ALL,<status_code>,<status_string>
<status_code> ::= an integer status
code
<status_string> ::= an ASCII status
string
n/a
:CALibrate:SWITch? (see
{PROTected | UNPRotected}
n/a
:CALibrate:TEMPerature?
(see
<return value> ::= degrees C delta
since last cal in NR3 format
n/a
:CALibrate:TIME? (see
<return value> ::=
<hours>,<minutes>,<seconds>; all in
NR1 format
Table 7
:CHANnel<n> Commands Summary
Command
Query
Options and Query Returns
:CHANnel<n>:BWLimit {{0
| OFF} | {1 | ON}} (see
:CHANnel<n>:BWLimit?
(see
{0 | 1}
<n> ::= 1-2 or 1-4 in NR1 format
:CHANnel<n>:COUPling
<coupling> (see
:CHANnel<n>:COUPling?
(see
<coupling> ::= {AC | DC}
<n> ::= 1-2 or 1-4 in NR1 format
:CHANnel<n>:DISPlay {{0
| OFF} | {1 | ON}} (see
:CHANnel<n>:DISPlay?
(see
{0 | 1}
<n> ::= 1-2 or 1-4 in NR1 format
:CHANnel<n>:IMPedance
<impedance> (see
:CHANnel<n>:IMPedance?
(see
<impedance> ::= {ONEMeg | FIFTy}
<n> ::= 1-2 or 1-4 in NR1 format
:CHANnel<n>:INVert {{0
| OFF} | {1 | ON}} (see
:CHANnel<n>:INVert?
(see
{0 | 1}
<n> ::= 1-2 or 1-4 in NR1 format
Summary of Contents for DSO6014A/L
Page 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Page 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Page 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Page 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Page 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Page 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...