398
Agilent 6000 Series Oscilloscopes Programmer's Reference
3
Commands by Subsystem
:TRIGger:IIC:TRIGger[:TYPE]
(see
Command Syntax
:TRIGger:IIC:TRIGger[:TYPE] <value>
<value> ::= {STARt | STOP | READ7 | READEeprom | WRITe7 | WRITe10
| NACKnowledge | ANACknowledge | R7Data2 | W7Data2 | RESTart}
The :TRIGger:IIC:TRIGger[:TYPE] command sets the IIC trigger type:
•
STARt — Start condition.
•
STOP — Stop condition.
•
READ7 — 7- bit address frame containing
(Start:Address7:Read:Ack:Data). The value READ is also accepted for
READ7.
•
R7Data2 — 7- bit address frame containing
(Start:Address7:Read:Ack:Data:Ack:Data2).
•
READEeprom — EEPROM data read.
•
WRITe7 — 7- bit address frame containing
(Start:Address7:Write:Ack:Data). The value WRITe is also accepted for
WRITe7.
•
W7Data2 — 7- bit address frame containing
(Start:Address7:Write:Ack:Data:Ack:Data2).
•
WRITe10 — 10- bit address frame containing (Start:Address
byte1:Write:Ack:Address byte 2:Data).
•
NACKnowledge — Missing acknowledge.
•
ANACknowledge — Address with no acknowledge.
•
RESTart — Another start condition occurs before a stop condition.
Query Syntax
:TRIGger:IIC:TRIGger[:TYPE]?
The :TRIGger:IIC:TRIGger[:TYPE]? query returns the current IIC trigger
type value.
Return Format
<value><NL>
<value> ::= {STAR | STOP | READ7 | READE | WRIT7 | WRIT10 | NACK | ANAC
| R7D2 | W7D2 | REST}
See Also
•
"Introduction to :TRIGger Commands"
•
N O T E
The short form of READ7 (READ7), READEeprom (READE), WRITe7 (WRIT7), and WRITe10
(WRIT10) do not follow the defined Long Form to Short Form Truncation Rules (see
Summary of Contents for DSO6014A/L
Page 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Page 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Page 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Page 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Page 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Page 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...