Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
237
:MARKer:X2Position
(see
Command Syntax
:MARKer:X2Position <position> [suffix]
<position> ::= X2 cursor position in NR3 format
<suffix> ::= {s | ms | us | ns | ps | Hz | kHz | MHz}
The :MARKer:X2Position command sets :MARKer:MODE to MANual, sets
the X2 cursor position and moves the X2 cursor to the specified value.
Query Syntax
:MARKer:X2Position?
The :MARKer:X2Position? query returns current X2 cursor position. If the
front- panel cursors are off an error is returned. This is functionally
equivalent to the obsolete :MEASure:TSTOp command/query.
Return Format
<position><NL>
<position> ::= X2 cursor position in NR3 format
See Also
•
"Introduction to :MARKer Commands"
•
•
•
•