Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
339
:TRIGger:NREJect
(see
Command Syntax
:TRIGger:NREJect <value>
<value> ::= {{0 | OFF} | {1 | ON}}
The :TRIGger:NREJect command turns the noise reject filter off and on.
When the noise reject filter is on, the trigger circuitry is less sensitive to
noise but may require a greater amplitude waveform to trigger the
oscilloscope. This command is not valid in TV trigger mode.
Query Syntax
:TRIGger:NREJect?
The :TRIGger:NREJect? query returns the current noise reject filter mode.
Return Format
<value><NL>
<value> ::= {0 | 1}
See Also
•