Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
173
:CHANnel<n>:IMPedance
(see
Command Syntax
:CHANnel<n>:IMPedance <impedance>
<impedance> ::= {ONEMeg | FIFTy}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :CHANnel<n>:IMPedance command selects the input impedance setting
for the specified analog channel. The legal values for this command are
ONEMeg (1 M
Ω
) and FIFTy (50
Ω
).
Query Syntax
:CHANnel<n>:IMPedance?
The :CHANnel<n>:IMPedance? query returns the current input impedance
setting for the specified channel.
Return Format
<impedance value><NL>
<impedance value> ::= {ONEM | FIFT}
See Also
•
"Introduction to :CHANnel<n> Commands"
N O T E
The analog channel input impedance of the 100 MHz bandwidth oscilloscope models is
fixed at ONEMeg (1 M
Ω
).