364
Agilent 6000 Series Oscilloscopes Programmer's Reference
3
Commands by Subsystem
:TRIGger[:EDGE] Commands
Table 66 :TRIGger[:EDGE] Commands Summary
Command
Query
Options and Query Returns
:TRIGger[:EDGE]:COUPlin
g {AC | DC | LF} (see
:TRIGger[:EDGE]:COUPlin
g? (see
{AC | DC | LF}
:TRIGger[:EDGE]:LEVel
<level> [,<source>]
(see
:TRIGger[:EDGE]:LEVel?
[<source>] (see
For internal triggers, <level> ::=
.75 x full-scale voltage from center
screen in NR3 format.
For external triggers, <level> ::= 2
volts with probe attenuation at 1:1
in NR3 format.
For digital channels (MSO models),
<level> ::= 8 V.
<source> ::= {CHANnel<n> | EXTernal}
for DSO models
<source> ::= {CHANnel<n> |
DIGital0,..,DIGital15 | EXTernal }
for MSO models
<n> ::= 1-2 or 1-4 in NR1 format
:TRIGger[:EDGE]:REJect
{OFF | LF | HF} (see
:TRIGger[:EDGE]:REJect?
(see
{OFF | LF | HF}
:TRIGger[:EDGE]:SLOPe
<polarity> (see
:TRIGger[:EDGE]:SLOPe?
(see
<polarity> ::= {POSitive | NEGative
| EITHer | ALTernate}
:TRIGger[:EDGE]:SOURce
<source> (see
)
:TRIGger[:EDGE]:SOURce?
(see
<source> ::= {CHANnel<n> | EXTernal}
for DSO models
<source> ::= {CHANnel<n> |
DIGital0,..,DIGital15 | EXTernal}
for MSO models
<n> ::= 1-2 or 1-4 in NR1 format
Summary of Contents for DSO6014A/L
Page 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Page 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Page 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Page 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Page 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Page 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...