Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
353
CAN Id specification is set by the :TRIGger:CAN:PATTern:ID
and:TRIGger:CAN:PATTern:ID:MODE commands.
CAN Data specification is set by the :TRIGger:CAN:PATTern:DATA
command.
CAN Data Length Code is set by the :TRIGger:CAN:PATTern:DATA:LENGth
command.
Query Syntax
:TRIGger:CAN:TRIGger?
The :TRIGger:CAN:TRIGger? query returns the current CAN trigger on
condition.
Return Format
<condition><NL>
<condition> ::= {SOF | DATA | ERR | IDD | IDE | IDR | ALL | OVER | ACK}
Errors
•
See Also
•
"Introduction to :TRIGger Commands"
•
•
•
":TRIGger:CAN:PATTern:DATA:LENGth"
•
•
":TRIGger:CAN:PATTern:ID:MODE"
•
":TRIGger:CAN:SIGNal:DEFinition"
•
N O T E
SOF is the only valid selection for analog oscilloscopes. If the automotive CAN and LIN
serial decode option (Option AMS) has not been licensed, SOF is the only valid selection.