:TRIGger:SEQuence:COUNt
(see
Command Syntax
:TRIGger:SEQuence:COUNt <count>
<count> ::= integer in NR1 format
The :TRIGger:SEQuence:COUNt command sets the sequencer edge counter
resource. The edge counter is used in the trigger stage to determine the
number of edges that must be found before the sequencer generates a
trigger.
Query Syntax
:TRIGger:SEQuence:COUNt?
The :TRIGger:SEQuence:COUNt? query returns the current sequencer edge
counter setting.
Return Format
<count><NL>
<count> ::= integer in NR1 format
See Also
•
"Introduction to :TRIGger Commands"
•
•