DMM Measurement Tests
You can set up a ±% variance threshold that is used to set
up the exception points on the chart.
Setting up DMM Tests
-1-
From the
Select Test Mode
screen, choose
Auto Tests
and press enter to display the
Select
Tests
screen.
-2-
Select
DMM test
and press
F2
Param
to display the
DMM Test Mode
screen.
Max DCV (±):
Enter the maximum ± DC Volts range
from 1 to 400.
Min DCV (±):
Enter the minimum ± DC Volts range
from 1 to 400.
Max Res:
Enter the maximum resistance range
from 0 to 1 MOhm.
Min Res:
Enter the minimum resistance range
from 0 to 1 MOhm.