Vacuum Gauge Controller AGC-100
78
Transmit:
ETH
[,a,bbb.bbb.bbb.bbb,ccc.ccc.ccc.ccc,ddd.ddd.ddd.ddd]
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,bbb.bbb.bbb.bbb,ccc.ccc.ccc.ccc,ddd.ddd.ddd.ddd <CR><LF>
Description
a
DHCP (dynamic host configuration protocol),
a =
0 –> Statically
1 –> Dynamically
bbb.bbb.bbb.bbb IP
address
ccc.ccc.ccc.ccc Subnet
address
ddd.ddd.ddd.ddd Gateway
address
"Transmit (T)" and "Receive (R)" are related to host.
T:
TID
<CR> [<LF>]
R: <ACK> <CR> <LF>
T: <ENQ>
R: PVG <CR> <LF>
Request for gauge identification
Positive acknowledgement
Request for data transmission
Gauge identification
T:
SP1
<CR> [<LF>]
R: <ACK> <CR> <LF>
T: <ENQ>
R: 1,1.0000E-09,9.0000E-07 <CR> <LF>
Request for parameters of switching function 1
(setpoint 1)
Positive acknowledgement
Request for data transmission
Thresholds
T:
SP1
,1,6.80E-3,9.80E-3 <CR> [<LF>]
R: <ACK> <CR> <LF>
Modification of parameters of switching
function 1 (setpoint 1)
Positive acknowledgement
T:
FOL
,2 <CR> [<LF>]
R: <NAK> <CR> <LF>
T: <ENQ>
R: 0001 <CR> <LF>
T:
FIL
,2 <CR> [<LF>]
R: <ACK> <CR> <LF>
T: <ENQ>
R: 2 <CR> <LF>
Modification of filter time constant (syntax error)
Negative acknowledgement
Request for data transmission
ERROR word
Modification of filter time constant
Positive acknowledgement
Request for data transmission
Filter time constants
T:
PR1
<CR> [<LF>]
R: <ACK> <CR> <LF>
T: <ENQ>
R: 0,8.3400E-03 <CR> <LF>
T: <ENQ>
R: 1,8.0000E-04 <CR> <LF>
Request for measurement data
Positive acknowledgement
Request for data transmission
Status and pressure
Request for data transmission
Status and pressure
ETH
- Ethernet configuration
5.13 Example
Summary of Contents for AGC-100
Page 2: ...Va AGC acuu C 100 um G Gaug ge Co Cop Ag ontro pyright 2016 gilent Inc oller ...
Page 3: ......
Page 24: ...Vacuum Gauge Controller AGC 100 24 Bargraph FSR ...
Page 92: ...Vacuum Gauge Controller AGC 100 92 Notes ...
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ...t qna96e1 ...