Vacuum Gauge Controller AGC-100
64
The full scale value of the measurement range (Full Scale) of linear
analog gauges has to be defined by the user. The full scale value of
linear digital gauges and logarithmic gauges is automatically recognized.
Transmit:
FSR
[,a] <CR>[<LF>]
Description
a
Full scale value, a =
0 –> 0.01 mbar
1 –> 0.01 Torr
2 –> 0.02 mbar
3 –> 0.02 Torr
4 –> 0.05 mbar
5 –> 0.05 Torr
6 –> 0.10 mbar
7 –> 0.10 Torr
8 –> 0.25 mbar
9 –> 0.25 Torr
10 –> 0.50 mbar
11 –> 0.50 Torr
12 –> 1 mbar
13 –> 1 Torr
14 –> 2 mbar
15 –> 2 Torr
16 –> 5 mbar
17 –> 5 Torr
18 –> 10 mbar
19 –> 10 Torr
20 –> 20 mbar
21 –> 20 Torr
22 –> 50 mbar
23 –> 50 Torr
24 –> 100 Torr
25 –> 100 mbar
26 –> 200 mbar
27 –> 200 Torr
28 –> 500 mbar
29 –> 500 Torr
30 –> 1000 mbar
31 –> 1100 mbar
32 –> 1000 Torr
33 –> 2 bar
34 –> 5 bar
35 –> 10 bar
36 –> 50 bar
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a
<CR><LF>
Description
a
Full scale value
FSR
- Measurement range
(linear gauges)