Vacuum Gauge Controller AGC-100
59
Transmit:
PR1
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,sx.xxxxEsxx
<CR><LF>
Description
a
Status, a =
0 –> Measurement data okay
1 –> Underrange
2 –> Overrange
3 –> Sensor error
4 –> Sensor off
5 –> No sensor
6 –> Identification error
7 –> Error FRG
sx.xxxxEsxx Measurement
value
1)
[in current pressure unit]
(s = sign)
1)
Values always in exponential format.
For logarithmic gauges, the 3
rd
and 4
th
decimal are always 0.
Transmit:
RES
[,a] <CR>[<LF>]
Description
a a
=
1 –> Cancels currently active error and re-
turns to measurement mode
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: b[,b][,b][…] <CR><LF>
Description
b
List of all present error messages, b =
0 –> No error
1 –> Watchdog has responded
2 –> Task fail error
3 –> FLASH error
4 –> RAM error
5 –> EEPROM error
6 –> DISPLAY error
7 –> A/D converter error
8 –> UART error
9 –> General error
10 –> ID error
PR1
- Measurement data
gauge
RES
- Reset
Summary of Contents for AGC-100