Vacuum Gauge Controller AGC-100
69
Transmit:
DCC
[,a] <CR>[<LF>]
Description
a
Contrast in percent, a = 0 … 100
100% = full contrast
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a
<CR><LF>
Description
a Contrast
Transmit:
DCS
[,a] <CR>[<LF>]
Description
a
Screensaver, a =
0 –> Off (default)
1 –> After 10 minutes
2 –> After 30 minutes
3 –> After 1 hour
4 –> After 2 hours
5 –> After 8 hours
6 –> The backlight is switched off completely
after 1 minute
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a
<CR><LF>
Description
a Screensaver
Transmit:
ERA
[,a] <CR>[<LF>]
Description
a
Switching behaviour error relay, a =
0 –> Switches for all errors (default)
1 –> Only unit errors
2 –> Error sensor and unit error
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a
<CR><LF>
Description
a
Switching behaviour error relay
DCC
- Display control contrast
DCS
- Display control
screensaver
ERA
- Error relay allocation