Vacuum Gauge Controller AGC-100
72
The group is only available when a USB memory stick formatted for the
the FAT file system (FAT32) is plugged in. Use a max. 32 GB memory
stick.
Transmit:
DAT
[,yyyy-mm-dd] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: yyyy-mm-dd
<CR><LF>
Description
yyyy-mm-dd
Current date in the format yyyy-mm-dd
Transmit:
LCM
[,a,b,c,ddddddd] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c,ddddddd
<CR><LF>
Description
a
Data logger command, a =
0 –> Stop / data logging stopped
1 –> Start / data logging started
2 –> Clear / deletion of measurement data file
(ending CSV) from USB memory stick
b
Data logging interval, b =
0 –> Logging interval 1/s
1 –> Logging interval 1/10 s
2 –> Logging interval 1/30 s
3 –> Logging interval 1/60 s
4 –> Logging interval in the event of measurement
value changes
≥
1%
5 –> Logging interval in the event of measurement
value changes
≥
5%
c
Decimal separator, c =
0 –> , (decimal comma)
1 –> . (decimal point)
ddddddd
File name (max. 7 digits)
Transmit:
TIM
[,hh:mm] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: hh:mm
<CR><LF>
Description
hh:mm
Current time in the format hh:mm [24 h]
5.9 Data Logger Parameters
DAT
- Date
LCM
- Start / stop data logger
TIM
- Time
Summary of Contents for AGC-100
Page 2: ...Va AGC acuu C 100 um G Gaug ge Co Cop Ag ontro pyright 2016 gilent Inc oller ...
Page 3: ......
Page 24: ...Vacuum Gauge Controller AGC 100 24 Bargraph FSR ...
Page 92: ...Vacuum Gauge Controller AGC 100 92 Notes ...
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ...t qna96e1 ...