Vacuum Gauge Controller AGC-100
36
The offset value is displayed and readjusted according to the actual measurement
value.
Available for the following gauges:
Pirani
(PVG)
Pirani / Capacitance
(PCG)
Cold cathode
(IMG)
Cold cathode / Pirani
(FRG-70x)
Hot ionization / Pirani
(FRG-720 / -730)
Capacitance
(CDG)
The offset correction affects:
the displayed measurement value
the displayed threshold value of the switching functions
the analog outputs at the
CONTROL
connector
(
→
20)
Wert
OFFSET
I
OFFSET OFF
I
Offset correction factory-deactivated
e.g.
OFFSET 9.53
I
Offset correction activated (display
in the relevant units of measure-
ment)
Press key for >1.5 s:
The offset value is readjusted (the actual measurement
value is accepted as new offset value).
Reset the offset value.
Save change and return to read mode.
When offset correction is activated, the saved offset value is subtracted from the
actual measurement value. This allows measuring relative to a reference pressure.
Offset correction of the
controller