Vacuum Gauge Controller AGC-100
62
Transmit:
DCD
[,a] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a
<CR><LF>
Description
a
Resolution a =
0 –> AUTO (default)
1 –> One digit
2 –> Two digits
3 –> Three digits
4 –> Four digits
When the PrE (
→
71) is ON and the pressure is in the range p<1.0E-4 mbar the
display resolution of the PCG75x gauges is reduced by one decimal digit.
Transmit:
DGS
[,a] <CR>[<LF>]
Description
a
Degas, a =
0 –> Degas off (default)
1 –> Degas on (3 minutes)
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a
<CR><LF>
Description
a Degas
status
Transmit:
EUM
[,a] <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a
<CR><LF>
Description
a
Emission, a =
0 –> Manually
1 –> Automatic (default)
DCD
- Display resolution
DGS
- Degas
EUM
- Emission user mode